Reported Nov 14, 2012 · Terminated
S5 Mast Roller Pump 85, Item Number: 10-88-60 The Stockert S5/Sorin C5 System is intended to be used during…
Sorin Group USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Sorin Group USA, Inc. · 14401 W 65th Way · Arvada, CO · 80004
Is this the product you have?
- S5 Mast Roller Pump 85
- Item Number10-88-60 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less
- Lot codes and dates
- Serial Numbers
50E50766--50E5076950E50772-50E5077450E50783-50E50784
- Serial Numbers
- Amount recalled
- 3 units
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
Why it was recalled
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Sorin Group USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.