Reported Nov 25, 2015 · Terminated
Prism Medical P-300 Portable Consumer Lift; PN 303050, 303050R, and 303051; weight specification 300 lbs. /136 kg.
ErgoSafe Products, LLC (DBA) Prism Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ErgoSafe Products, LLC (DBA) Prism Medical · 10888 Metro Ct · Maryland Heights, MO · 63043-2413
Is this the product you have?
Prism Medical P-300 Portable Consumer LiftPN 303050303050Rand 303051weight specification 300 lbs. /136 kg
- Lot codes and dates
- All serial numbers produced since October 2011
- Part Number 303050SN
1,098 of 1,098 codes
P303885P305001P303816P303948P304050P304745P303670P303783P303822P304003P303999P303675P303342P303351P303519P303690P303784P304045P303405P305150P303741P303825P303916P304732P303595P303850P304597P304606P305056P305095P303547P304715P305040P304598P303458P303945P304012P304016P304438P304740P303528P303552P303684P304705P303802P303820P303658P305099P304707P303051P303106P303484P303487P303688P303756P303771P303863P303866P303870P303600
- CQ303188 CQ303194 Part Number 303050RSN
P303347P303353P303453P303575P303457P303531P303067P303073P303401P303077P303343P303076P303091P303543P303074P303052
- P303577 P303498 Part Number 303051SN
448 of 448 codes
P304182P304184P304488P304604P304911P304409P304417P304450P304063P303856P304522P304406P304413P304424P304447P304519P304551P304977P304064P304425P305154P305070P305023P305110P305022P304407P304418P305064P304863P304412C303033C303091C303097C303137C303139C303144C303146C303148C303151C303162C303189C303193P305174P304916P304189P304148P305024P304554P304877P305019P305063P304120P304179P304423P304912P305131P304602P304069P304188P304542
- Amount recalled
- 1,721 lifts
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Drive shaft and over speed arm may simultaneously fail potentially resulting in a patient falling during use
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new ErgoSafe Products, LLC (DBA) Prism Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.