Reported Nov 21, 2012 · Terminated
PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor…
Dako North America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Dako North America Inc. · 6392 Via Real · Carpinteria, CA · 93013-2921
Is this the product you have?
- PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.
- Lot codes and dates
- Catalog #: PT-Module
- Amount recalled
- 136 units
Where it was sold
Why it was recalled
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
This product could cause temporary or reversible health problems — stop using it.