Reported Nov 7, 2018 · Terminated

Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is…

Stryker GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker GmbH · Bohnackerweg 1 · Selzach, N/A · N/A

Is this the product you have?

  • Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults
  • for the treatment and fixation ofOpen and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity Correction of segmental bony or soft tissue defects Joint arthrodesis Management of communicated intra-articular fractures of the distal radius Bone transport The Hoffmann
  • LRF System is indicated in adults forOsteotomy Revision procedure where other treatments or devices have been unsuccessful Bone reconstruction procedures Fusions and replantations of the foot Charcot foot reconstruction Lisfranc dislocations The typical LRF construct features two or more rings bridged by three to four telescopic struts with multiplanar, ball jointed fixation bolts. Once applied, the LRF can be unlocked for gross frame adjustment and fracture reduction. The LRF can be finely adjusted to aid in limb alignment and definitively locked to hold compression and reduction. If deemed appropriate, the LRF Telescopic Struts can provide up to 5mm of controlled dynamization. Original Statement OP-Tech H-ST-1_Rev2_HoffmannLRF_OpTech_2016-10690]
Lot codes and dates
  • Lots
    • L18238
    • L13689
    • L22249
    • L26333
    • L31075
    • L31035
    • L35849
    • L35850
    • L45970
    • L46580
    • L47080
    • L46577
    • L46781
    • L47195
    • L49918
    • L49816
    • L51362
    • L53106
    • L53487
    • L53823
    • L53488
    • L56830
    • L56831
    • L57502
    • Y00081
    • Y00080
    • Y02872
    • Y02873
    • Y03128
    • Y07628
    • Y07630
    • Y08561
    • Y08555
    • Y08554
    • Y08553
    • Y12876
    • Y14867
    • Y14641
Amount recalled
1355

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

Why it was recalled

The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 4, 2021 · Data as of Sep 19, 5:34 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.