Reported Nov 20, 2024 · Ongoing

Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100

Boston Scientific Corporation

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Boston Scientific Corporation · 4100 Hamline Ave N · Saint Paul, MN · 55112-5700

Is this the product you have?

  • Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100
Lot codes and dates
  • GTIN 08714729992660
Amount recalled
20290 units

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Why it was recalled

Stop using this product now

Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Boston Scientific Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:43 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.