Reported Nov 2, 2016 · Terminated
ReTurn 7500/7500 Aid, transfer
Handicare USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Handicare USA, Inc. · 2201 Hangar Pl · Allentown, PA · 18109-9342
Is this the product you have?
- ReTurn 7500/7500 Aid, transfer
- Lot codes and dates
- Model # 7400, 7500 and 7322 (spare part) Serial # 373317697270501334, 373317697270501430, 073317697400500012, 073317697400500057
- Amount recalled
- 29 units , plus 47 wing handles were shipped as spare parts
Where it was sold
Why it was recalled
During the period of April 2009 May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.