Reported Dec 1, 2021 · Ongoing
BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660…
Becton Dickinson Infusion Therapy Systems Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson Infusion Therapy Systems Inc. · 9450 S State St · Sandy, UT · 84070-3213
Is this the product you have?
- BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.
- Lot codes and dates
3 of 3 products
Products covered by this recall # Barcodes Product as published 1 Catalog number/Model number: 383536 2 50382903835367Batch numbers/UDI numbers: Lot: 1193055/(17)240630(10)1193055(30)80/(01)50382903835367, 2024-06-30 3 503829038353675038290383536750382903835367Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1188953: (17)240630(10)1188953(30)80/(01)50382903835367, 2024-06-30 - Amount recalled
- 155,840 devices, 80 units in each box
Where it was sold
Why it was recalled
There is a breach in the product packaging that renders the product non-sterile.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Becton Dickinson Infusion Therapy Systems Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.