Reported Nov 9, 2016 · Terminated
Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200,…
Hill-Rom, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hill-Rom, Inc. · 1069 State Road 46 E · Batesville, IN · 47006-7520
Is this the product you have?
Product Number 3102071 Ceiling BracketArch 71 White with
- use on ceiling installed rail systems
Multirall 200Likorall 200242243250and LikoGuard L and XL
- Lot codes and dates
manufacturing date between Dec 012015- Apr 142016
- Amount recalled
- 188
Where it was sold
Why it was recalled
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
This product could cause temporary or reversible health problems — stop using it.