Reported Dec 15, 2021 · Ongoing
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information…
Covidien LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Covidien Llc · 15 Hampshire St · Mansfield, MA · 02048-1113
Is this the product you have?
- Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress
- and battery pack status Item CodeCBCA
- Lot codes and dates
- Serial NumbersMCSB006939
- MCSB006947 GTIN10884521592896
- Amount recalled
- 2 units OUS
This recall covers 17 products
The agency filed one notice per product. You are reading one of them; the other 16 are below.
4 more are in the archive.
Why it was recalled
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Covidien LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.