Reported Dec 9, 2015 · Terminated
Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are…
Stryker Medical Division of Stryker Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Stryker Medical Division of Stryker Corporation · 3800 E Centre Ave · Portage, MI · 49002-5826
Is this the product you have?
- Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following parent
- devicesModel 2030 (EPIC II), model 3005 S3 beds and stretcher. models 1105, 1105E, 1105X, 1105XE, 1115, 1115E, 1115X, 1115XE, 1125E, 1125X and 1125XE
- Lot codes and dates
- Stryker 2-Stage and 3-Stage I.V. Poles distributed on beds, stretchers, and in service kits. This product field action is limited to I.V. Poles distributed between 01/07/2014 - 06/24/2014. The following I.V. Pole option numbers
- come affixed to beds and stretchersManufacturer Model Number and Product Description 2035-112-010; Head End Left, 2 Stage I.V. Pole 2035-113-011; Head End Right, 2 Stage I.V. Pole 0785-035-101; 26", 2 Stage IV Pole Assembly 0785-035-200; 26", 3 Stage IV Pole Assembly 0785-035-300; 30", 3 Stage IV Pole Assembly 0785-035-401; 30", 2 Stage IV Pole Assembly The following I.V. Pole option numbers ship as service kits. Manufacturer Service Kit Number and Product Description 1105-035-368; 30" IV POLE 2 STAGE HE PR 1105-035-367; 30" IV POLE 2 STAGE HE PL 1105-035-343; 30" IV POLE 2 STAGE FE PR 1105-035-341; 30" IV POLE, 2 STAGE HE, PR 1105-035-340; 30" IV POLE, 2 STAGE FE, PL 1105-035-338; 30" IV POLE, 2 STAGE FE, PL 1040-035-338; 2-STAGE IV POLEHEAD LT30" ZM 0785-035-340; 2-STAGE IV POLE,HEAD LT, 30" 0785-035-341; 2-STAGE IV POLEHEAD RT 30" 0785-035-343; 2-STAGE IV POLE,FOOT RT, 30" 1105-035-344; 30" IV POLE, 3 STAGE FE, PL 1105-035-342; 30" IV POLE, 3 STAGE HE, PL 1105-035-339; 30" IV POLE, 3 STAGE FE, PR 1105-035-369; 30" IV POLE, 3 STAGE HE, PL 1105-035-366; 30" IV POLE, 3 STAGE HE, PR 0785-035-339; 3-STAGE IV POLE,FOOT RT, 30" 0785-035-342; 3-STAGE IV POLE,HEAD LT, 30" 1105-035-644; 26" IV POLE, 3 STAGE FL 1105-035-642; 26" IV POLE, 3 STAGE HL 1105-035-639; 26" IV POLE 3 STAGE FR 1105-035-637; 26" IV POLE, 3 STAGE HR 0785-035-637; 3-STAGE IV POLE,HEAD RT, 26" 0785-035-639; 3-STAGE IV POLE,FOOT RT, 26" 0785-035-642; 3-STAGE IV POLE,HEAD LT, 26" 1105-035-643; 26" IV POLE, 2 STAGE FR 1105-035-641; 26" IV POLE, 2 STAGE HR 1105-035-640; 26" IV POLE, 2 STAGE FL 1105-035-638; 26" IV POLE, 2 STAGE HL 0785-035-638; 2-STAGE IV POLE,HEAD LT, 26" 0785-035-640; 2-STAGE IV POLE,FOOT LT, 26" 0785-035-641; 2-STAGE IV POLE,HEAD RT, 26" 0785-035-643; 2-STAGE IV POLE,FOOT RT, 26" 2035-112-000; IV POLE - 2 STAGE H/E 2035-113-000; DUAL H/E 2 STAGE IV POLE OPT
- Amount recalled
- 19,876
Where it was sold
- Nationwide
- DC
Why it was recalled
It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Stryker Medical Division recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.