Reported Nov 20, 2019 · Ongoing
TriMed Peg Guide Extender, REF PEG-XTNDR, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.
TriMed Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: TriMed Inc. · 27533 Avenue Hopkins · Santa Clarita, CA · 91355-3910
Is this the product you have?
- TriMed Peg Guide Extender, REF PEG-XTNDR, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.
- Lot codes and dates
Lot number 28582861and 3749
- Amount recalled
- 1,903 Peg Guide Extenders
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
The spring on the PEG-Extender can rust due to the type of steel used to manufacture it.