Reported Nov 29, 2023 · Ongoing
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007,…
Instrumentation Laboratory
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Instrumentation Laboratory · 180 Hartwell Rd · Bedford, MA · 01730-2443
Is this the product you have?
GEM Premier 3000 PAK - pHpCO2pO2HctNa+K+Ca++Part Numbers 0002430750700024315007000243300070002434500700024360007
- Lot codes and dates
5 of 5 products
Products covered by this recall Model / number Barcodes Product as published 0002430750708426950087946UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C 0002431500708426950082484UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C 0002433000708426950082514UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J 0002434500708426950082545UDI-DI: 08426950082545, Lot Numbers: 323761H 0002436000708426950082569UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E - Amount recalled
- 2448 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Instrumentation Laboratory recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.