Reported Dec 9, 2015 · Terminated

Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and…

ConMed Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ConMed Corporation · 525 French Rd · Utica, NY · 13502-5945

Is this the product you have?

  • Altrus Thermal Tissue Fusion Energy Sources
  • Catalog Numbers1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source
Lot codes and dates
  • Serial Numbers
    • 1) 15090103 & 15150005
    • 15150006
    • 15150007
    • 15150008
    • 15150009
    • 15150010
    • 15150011
    • 15150012
    • 15130004
    • 15250013
    • 15250014
    • 15250015
    • 15250016
    • 15250017
    • 15300018 & 15300020
    • 2) 15050001
    • 15300005 & 15300006
Amount recalled
9 devices

Where it was sold

Why it was recalled

Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new ConMed Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 24, 2018 · Data as of Sep 19, 5:59 PM UTC