Reported Nov 27, 2013 · Terminated

Salter Labs EtCO2 Sample line 15' (4.6 M) tube, male luer connectors. Rx only; Made in Mexico; Manufacturer:Salter…

Salter Labs

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: SALTER LABS · 100 Sycamore Rd · Arvin, CA · 93203-2300

Is this the product you have?

  • Salter Labs EtCO2 Sample line 15' (4.6 M) tube, male luer connectors. Rx only; Made in Mexico
  • ManufacturerSalter Labs, 100 West Sycamore Road, Arvin, CA 93203. CO2 sample line
Lot codes and dates
  • Model number
    • 4515-15-25
    • lot numbers 071513
    • 071513A and 120312
Amount recalled
91 cases (25 units per case)

Where it was sold

Why it was recalled

It has been determined that three lots of model number 4515-15-25 CO2 Sample Line 15' tube, male Luer lock connectors may contain a 30' tube instead of the 15' length.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Salter Labs recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 17, 2013 · Data as of Sep 19, 6:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.