Reported Nov 20, 2019 · Terminated

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product…

Tec Com GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Tec Com Gmbh · Eingang B · Kabelsketal, N/A · N/A

Is this the product you have?

    • TECOTHERM NEO
    • 100-130/200-240 VAC
    • 50-60 Hz/ max. 350VA
    • Class I
    • Schutzklasse I IP 20
  • UDI04260498580002 -
  • Product UsageThe Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective
Lot codes and dates
  • Serial Numbers
    • 2018/09/07
    • 2018/10/01
    • 2018/10/02
    • 2018/10/03
    • 2018/10/05
Amount recalled
5 units

Where it was sold

  • Nationwide
  • UT

Why it was recalled

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Tec Com GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 20, 2019 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.