Reported Nov 20, 2024 · Ongoing
Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general…
Stryker Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Stryker Corporation · 1941 Stryker Way · Portage, MI · 49002-9711
Is this the product you have?
- Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system
- Catalog Number0703-046-000
- Lot codes and dates
- GTIN07613327117202
- Lot Numbers
205 of 205 codes
544097785969216862016967632926546432420064855491659619076732542857699286603135436201697863292655643242016485549865961908673775035773832960313544622966666363284764324207648554996596190967481708577383306031354562296667636328486432420864855500660129176748170957738331603135466229666863632851643548446485550166012918674817105773833260313547623821036363285264354845649732486613341267481711577383336031354962382104638426166435484664973249663873446748171257738334608341196238210563842618
- Amount recalled
- 1,996,010 eaches
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Stryker Corporation recalls