Reported Nov 16, 2016 · Terminated

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human…

Abbott Laboratories

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Laboratories · 100 Abbott Park Rd · Abbott Park, IL · 60064-3502

Is this the product you have?

  • ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.
Lot codes and dates
  • Lot31497UN14 Exp.: 17 Oct. 2016
  • Lot09353UNI15 Exp.: 12 Feb. 2017
  • Lot45216UN15 12 May 2017
  • Lot14596UN15 Exp.: 30 Jul. 2017
  • Lot37055UN15 Exp.: 12 Oct. 2017
  • Lot14782UN16 Exp.: 21 Apr. 2018
Amount recalled
26, 837 kits

Where it was sold

  • Nationwide

Why it was recalled

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abbott Laboratories recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 17, 2017 · Data as of Sep 19, 4:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.