Reported Nov 20, 2019 · Terminated
ORTHOFIX Catalogue Number: ref 1-1100201,DRILL BIT Diameter .4,8 MM L.240 MM TIN COATED - QUICK CONNECT, RX ONLY, Non…
Orthofix Srl
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Orthofix Srl · Via Delle Nazioni 9 O 73 · Veronaverona, N/A · N/A
Is this the product you have?
- ORTHOFIX Catalogue Number
ref 1-1100201DRILL BIT Diameter .48 MM L.240 MM TIN COATED - QUICK CONNECTRX ONLYNon Sterile
- UDI(01)18033509856068
- Lot codes and dates
Lot # B1205238B1205760B1213842289814300239289814273343
- Amount recalled
- N/A
This recall covers 31 products
The agency filed one notice per product. You are reading one of them; the other 30 are below.
18 more are in the archive.
Where it was sold
Why it was recalled
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Orthofix Srl recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.