Reported Nov 14, 2018 · Terminated

TB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right…

Oscor, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Oscor, Inc. · 3816 Desoto Blvd · Palm Harbor, FL · 34683-1618

Is this the product you have?

  • TB Series Bipolar Pacing Leads, French Size 4F, 5F, 6F for Curve Type Straight, Atrial J, 60 degree Curve, and Right Heart. TB is packaged for sterile single use in a double tray. The lead is placed in a PETG inner tray that is sealed with a Tyvek inner lid. The sealed inner tray is placed in an outer PETG tray. The outer tray is sealed with an outer Tyvek lid. An inner label is placed on the outer Tyvek lid. The sealed trays are placed in a labeled white shelf box with on Instruction for use. Packed on lead per outer white shelf box.
Lot codes and dates

21 of 21 products

Products covered by this recall
Model / numberProduct as published
020004TB Lead, 4F Un-Shrouded, - 2B-81-532Z-D (OR-03868), 2B-81-532Z-D (OR-03874), 2B-81-532Z-D (C1-06951), 2B-81-532Z-D (C1-06993), 2B-81-532Z-D (C1-06994), 2B-81-532Z-D (C1-07016), 2B-81-532Z-D (C1-07018), 2B-81-532Z-D (C1-07017), 2B-81-532Z-D (D1-02633), 2B-81-532Z-D (D1-02883), 2B-81-532Z-D (DP-01395), 2B-81-532Z-D (OR-04404), 2B-81-532Z-D (C1-09074), 2B-81-532Z-D (C1-09166), 2B-81-532Z-D (DP-02801), 2B-81-532Z-D (DP 02801), 2B-81-532Z-D (DP-04451), 2B-81-532Z-D (DP-05052), 2B-81-532Z-D (DP-05115), 2B-81-532Z-D (DP-05225), 2B-81-532Z-D (DP-05371), 2B-81-532Z-D (DP-05482), 2B-81-532Z-D (DP-05516), 2B-81-532Z-D (DP-05938), 2B-81-532Z-D (DP05968), 2B-81-532Z-D (DP-05976), 2B-81-532Z-D (DP-05998), 2B-81-532Z-D (DP-06062), 2B-81-532Z-D (DP-06098), 2B-81-532Z-D (DP-06203), 2B-81-532Z-D (DP-07789), 2B-81-532Z-D (DP-07840), 2B-81-532Z-D (DP-08130)
020005TB LEAD, 5F UN-SHROUDED, - 2B-81-532Z-E (D2-01795), 2B-81-532Z-E (C1-06923), 2B-81-532Z-E (C1-06946), 2B-81-532Z-E (C1-06947), 2B-81-532Z-E (C1-06945), 2B-81-532Z-E (C1-06997), 2B-81-532Z-E (C1-06995), 2B-81-532Z-E (C1-06996), 2B-81-532Z-E (C1-07013), 2B-81-532Z-E (C1-07015), 2B-81-532Z-E (C1-07012), 2B-81-532Z-E (C1-07014), 2B-81-532Z-E (D2-01939), 2B-81-532Z-E (D1-02542), 2B-81-532Z-E (D1-02805), 2B-81-532Z-E (DP-01259), 2B-81-532Z-E (DP-01455), 2B-81-532Z-E (DP-01554), 2B-81-532Z-E (OR-04380), 2B-81-532Z-E (C1-08496), 2B-81-532Z-E (C1-08486), 2B-81-532Z-E (C1-08497), 2B-81-532Z-E (C1-08495), 2B-81-532Z-E (OR-04407), 2B-81-532Z-E (C1-08487), 2B-81-532Z-E (C1-08488), 2B-81-532Z-E (C1-08499), 2B-81-532Z-E (C1-08494), 2B-81-532Z-E (C1-08498), 2B-81-532Z-E (C1-08678), 2B-81-532Z-E (C1-08676), 2B-81-532Z-E (C1-08677), 2B-81-532Z-E (C1-08686), 2B-81-532Z-E (C1-08687), 2B-81-532Z-E (C1-08688), 2B-81-532Z-E (C1-08689), 2B-81-532Z-E (C1-08691), 2B-81-532Z-E (C1-08685), 2B-81-532Z-E (C1-08679), 2B-81-532Z-E (C1-08693), 2B-81-532Z-E (C1-08694), 2B-81-532Z-E (DP-02161), 2B-81-532Z-E (C1-08692), 2B-81-532Z-E (C1-09046), 2B-81-532Z-E (C1-09047), 2B-81-532Z-E (C1-09048), 2B-81-532Z-E (C1-09251), 2B-81-532Z-E (C1-09252), 2B-81-532Z-E (C1-09253), 2B-81-532Z-E (DP-02766), 2B-81-532Z-E (DP-02790), 2B-81-532Z-E (DP-02844), 2B-81-532Z-E (DP-03289), 2B-81-532Z-E (DP-03247), 2B-81-532Z-E (DP-03424), 2B-81-532Z-E (DP-04422), 2B-81-532Z-E (DP-04455), 2B-81-532Z-E (DP-04504), 2B-81-532Z-E (DP-04772), 2B-81-532Z-E (DP-04815), 2B-81-532Z-E (DP-04817), 2B-81-532Z-E (DP-04957), 2B-81-532Z-E (DP-05027), 2B-81-532Z-E (DP-05044), 2B-81-532Z-E (DP-05045), 2B-81-532Z-E (DP-05116), 2B-81-532Z-E (DP-05045_1), 2B-81-532Z-E (DP 05116), 2B-81-532Z-E (DP-05274), 2B-81-532Z-E (DP-05417), 2B-81-532Z-E (DP-05543), 2B-81-532Z-E (DP-05686), 2B-81-532Z-E (DP-05687), 2B-81-532Z-E (DP-05958), 2B-81-532Z-E (DP-06080), 2B-81-532Z-E (DP-06060), 2B-81-532Z-E (DP-06141), 2B-81-532Z-E (DP-06206), 2B-81-532Z-E (DP-06346), 2B-81-532Z-E (DP-06535), 2B-81-532Z-E (DP-06762), 2B-81-532Z-E (DP-06568), 2B-81-532Z-E (DP-07015), 2B-81-532Z-E (DP-07018), 2B-81-532Z-E (DP-07017), 2B-81-532Z-E (DP-07146), 2B-81-532Z-E (DP-07262), 2B-81-532Z-E (DP-07438), 2B-81-532Z-E (DP-07500), 2B-81-532Z-E (DP-07642), 2B-81-532Z-E (DP-07787), 2B-81-532Z-E (DP-07643)
TB LEAD, 6F UN-SHROUDED, 020006 - 2B-81-532Z-F (D2-01775), 2B-81-532Z-F (D2-01691), 2B-81-532Z-F (OR-03853), 2B-81-532Z-F (OR-03961), 2B-81-532Z-F (OR-03878), 2B-81-532Z-F (OR-03939), 2B-81-532Z-F (C1-06953), 2B-81-532Z-F (C1-06954), 2B-81-532Z-F (C1-06955), 2B-81-532Z-F (C1-06956), 2B-81-532Z-F (OR-04060), 2B-81-532Z-F (D1-02562), 2B-81-532Z-F (OR-04100), 2B-81-532Z-F (C1-07081), 2B-81-532Z-F (C1-07082), 2B-81-532Z-F (C1-07079), 2B-81-532Z-F (C1-07080), 2B-81-532Z-F (D1-02806), 2B-81-532Z-F (DP-01003), 2B-81-532Z-F (DP-01004), 2B-81-532Z-F (DP-01214), 2B-81-532Z-F (C1-07955), 2B-81-532Z-F (DP-01082), 2B-81-532Z-F (C1-07945), 2B-81-532Z-F (C1-07946), 2B-81-532Z-F (C1-07958), 2B-81-532Z-F (C1-07964), 2B-81-532Z-F (C1-07956), 2B-81-532Z-F (OR-04278), 2B-81-532Z-F (C1-07966), 2B-81-532Z-F (DP-01569), 2B-81-532Z-F (C1-07957), 2B-81-532Z-F (C1-07959), 2B-81-532Z-F (C1-07960), 2B-81-532Z-F (C1-07961), 2B-81-532Z-F (C1-07962), 2B-81-532Z-F (C1-07963), 2B-81-532Z-F (C1-07965), 2B-81-532Z-F (DP-01374), 2B-81-532Z-F (DP-01492), 2B-81-532Z-F (DP-01553), 2B-81-532Z-F (OR-04329), 2B-81-532Z-F (OR-04330), 2B-81-532Z-F (DP-01601), 2B-81-532Z-F (C1-08396), 2B-81-532Z-F (C1-08407), 2B-81-532Z-F (C1-08410), 2B-81-532Z-F (C1-08409), 2B-81-532Z-F (C1-08411), 2B-81-532Z-F (C1-08412), 2B-81-532Z-F (OR-04367), 2B-81-532Z-F (C1-08398), 2B-81-532Z-F (OR-04408), 2B-81-532Z-F (OR-04369), 2B-81-532Z-F (OR-04399), 2B-81-532Z-F (OR-04368), 2B-81-532Z-F (OR-04366), 2B-81-532Z-F (C1-08670), 2B-81-532Z-F (C1-08666), 2B-81-532Z-F (C1-08667), 2B-81-532Z-F (C1-08668), 2B-81-532Z-F (C1-08669), 2B-81-532Z-F (C1-08674), 2B-81-532Z-F (C1-08675), 2B-81-532Z-F (C1-08704), 2B-81-532Z-F (C1-08705), 2B-81-532Z-F (C1-08706), 2B-81-532Z-F (C1-08711), 2B-81-532Z-F (C1-08718), 2B-81-532Z-F (C1-08720), 2B-81-532Z-F (C1-08709), 2B-81-532Z-F (C1-08735), 2B-81-532Z-F (C1-08732), 2B-81-532Z-F (C1-08710), 2B-81-532Z-F (C1-08714), 2B-81-532Z-F (C1-08717), 2B-81-532Z-F (C1-08716), 2B-81-532Z-F (C1-08721), 2B-81-532Z-F (C1-08719), 2B-81-532Z-F (C1-08703), 2B-81-532Z-F (C1-08707), 2B-81-532Z-F (C1-08708), 2B-81-532Z-F (C1-08733), 2B-81-532Z-F (C1-08734), 2B-81-532Z-F (C1-08731), 2B-81-532Z-F (C1-08715), 2B-81-532Z-F (C1-08671), 2B-81-532Z-F (C1-08713), 2B-81-532Z-F (C1-08680), 2B-81-532Z-F (C1-08755), 2B-81-532Z-F (C1-08681), 2B-81-532Z-F (C1-08683), 2B-81-532Z-F (C1-08701), 2B-81-532Z-F (C1-08723), 2B-81-532Z-F (C1-08672), 2B-81-532Z-F (C1-08673), 2B-81-532Z-F (C1-08682), 2B-81-532Z-F (C1-08702), 2B-81-532Z-F (C1-08712), 2B-81-532Z-F (C1-08722), 2B-81-532Z-F (C1-08724), 2B-81-532Z-F (C1-08753), 2B-81-532Z-F (C1-08754), 2B-81-532Z-F (C1-08756), 2B-81-532Z-F (C1-08757), 2B-81-532Z-F (C1-08759), 2B-81-532Z-F (C1-08760), 2B-81-532Z-F (C1-08761), 2B-81-532Z-F (C1-08762), 2B-81-532Z-F (C1-08763), 2B-81-532Z-F (C1-08758), 2B-81-532Z-F (DP-02162), 2B-81-532Z-F (DP-02793), 2B-81-532Z-F (DP-02932), 2B-81-532Z-F (DP-02942), 2B-81-532Z-F (C1-09444), 2B-81-532Z-F (DP-03114), 2B-81-532Z-F (DP-03116), 2B-81-532Z-F (DP-03117), 2B-81-532Z-F (DP-03118), 2B-81-532Z-F (DP-03115), 2B-81-532Z-F (DP-03144), 2B-81-532Z-F (DP-03123), 2B-81-532Z-F (DP-03229), 2B-81-532Z-F (DP-03323), 2B-81-532Z-F (OR-04667), 2B-81-532Z-F (DP-03893), 2B-81-532Z-F (DP-03930), 2B-81-532Z-F (DP-04378), 2B-81-532Z-F (DP-04377), 2B-81-532Z-F (DP-04386), 2B-81-532Z-F (DP-04454), 2B-81-532Z-F (DP-04456), 2B-81-532Z-F (DP-04707), 2B-81-532Z-F (DP-04760), 2B-81-532Z-F (DP-04818), 2B-81-532Z-F (DP-05108), 2B-81-532Z-F (DP-05362), 2B-81-532Z-F (DP-05349), 2B-81-532Z-F (DP-05350), 2B-81-532Z-F (DP-05416), 2B-81-532Z-F (DP-05510), 2B-81-532Z-F (DP-05544), 2B-81-532Z-F (DP-05559), 2B-81-532Z-F (DP-05511), 2B-81-532Z-F (DP-05599), 2B-81-532Z-F (DP-05630), 2B-81-532Z-F (DP-05637), 2B-81-532Z-F (DP-05685), 2B-81-532Z-F (DP-06005), 2B-81-532Z-F (DP-06052), 2B-81-532Z-F (DP-06388), 2B-81-532Z-F (DP-06495), 2B-81-532Z-F (DP-06496), 2B-81-532Z-F (DP-06497), 2B-81-532Z-F (DP-06498), 2B-81-532Z-F (DP-06500), 2B-81-532Z-F (DP-06546), 2B-81-532Z-F (DP-06501), 2B-81-532Z-F (DP-06567), 2B-81-532Z-F (DP-07049), 2B-81-532Z-F (DP-07134), 2B-81-532Z-F (DP-07513), 2B-81-532Z-F (DP-07784), 2B-81-532Z-F (DP-07785), 2B-81-532Z-F (DP-07790), 2B-81-532Z-F (DP-08166), 2B-81-532Z-F (DP-08167)
TB LEAD, 4F UN-SHROUDED ATRIAL J-TIP, 020010 - 2B-81-532Z-J (C1-06957), 2B-81-532Z-J (DP-01059), 2B-81-532Z-J (DP-04648)
TB LEAD, 5F UN-SHROUDED ATRIAL J-TIP, 020011 - 2B-81-532Z-K (C1-06940), 2B-81-532Z-K (D1-02916), 2B-81-532Z-K (D1-02559), 2B-81-532Z-K (OR-04406), 2B-81-532Z-K (DP-04647)
TB LEAD, 6F UN-SHROUDED ATRIAL J-TIP, 020012 - 2B-81-532Z-L (D1-02565), 2B-81-532Z-L (OR-04370), 2B-81-532Z-L (DP-04649)
TB LEAD, 4F UN-SHROUDED 60 DEG CURVE, 020016 - 2B-81-532Z-P (C1-06952), 2B-81-532Z-P (DP-01058)
TB LEAD, 5F UN-SHROUDED 60 DEG CURVE, 020017 - 2B-81-532Z-Q (OR-04416), 2B-81-532Z-Q (DP-03054)
TB LEAD, 6F UN-SHROUDED 60 DEG CURVE, 020018 - 2B-81-532Z-R (D2-01777), 2B-81-532Z-R (OR-03879), 2B-81-532Z-R (D1-02490), 2B-81-532Z-R (D2-01991), 2B-81-532Z-R (D1-02567), 2B-81-532Z-R (DP-01105), 2B-81-532Z-R (DP-01359), 2B-81-532Z-R (OR-04277), 2B-81-532Z-R (OR-04280), 2B-81-532Z-R (DP-01357), 2B-81-532Z-R (DP-01490), 2B-81-532Z-R (OR-04371), 2B-81-532Z-R (OR-04383), 2B-81-532Z-R (DP-06207)
TB LEAD, 4F UNSHROUDED RIGHT HEART CURVE, 020022 - 2B-81-532Z-V (DP-02185), 2B-81-532Z-V (DP-03321), 2B-81-532Z-V (DP-04708), 2B-81-532Z-V (DP-04709), 2B-81-532Z-V (DP-05354), 2B-81-532Z-V (DP-05415), 2B-81-532Z-V (DP-05414), 2B-81-532Z-V (DP-06061), 2B-81-532Z-V (DP-06760), 2B-81-532Z-V (DP-07891)
TB LEAD, 5F UN-SHROUDED RIGHT HEART CURVE, 020023 - 2B-81-532Z-W (D1-02822), 2B-81-532Z-W (D1-02817), 2B-81-532Z-W (D1-02924), 2B-81-532Z-W (D1 02817), 2B-81-532Z-W (C1-08091), 2B-81-532Z-W (C1-08096), 2B-81-532Z-W (C1-08098), 2B-81-532Z-W (C1-08095), 2B-81-532Z-W (C1-08097), 2B-81-532Z-W (C1-08133), 2B-81-532Z-W (C1-08134), 2B-81-532Z-W (DP-01116), 2B-81-532Z-W (C1-08135), 2B-81-532Z-W (C1-08101), 2B-81-532Z-W (C1-08121), 2B-81-532Z-W (C1-08124), 2B-81-532Z-W (C1-08123), 2B-81-532Z-W (C1-08128), 2B-81-532Z-W (C1-08137), 2B-81-532Z-W (C1-08099), 2B-81-532Z-W (C1-08103), 2B-81-532Z-W (C1-08104), 2B-81-532Z-W (C1-08106), 2B-81-532Z-W (C1-08112), 2B-81-532Z-W (C1-08113), 2B-81-532Z-W (C1-08114), 2B-81-532Z-W (C1-08115), 2B-81-532Z-W (C1-08117), 2B-81-532Z-W (C1-08118), 2B-81-532Z-W (C1-08120), 2B-81-532Z-W (C1-08126), 2B-81-532Z-W (C1-08127), 2B-81-532Z-W (C1-08136), 2B-81-532Z-W (C1-08092), 2B-81-532Z-W (C1-08100), 2B-81-532Z-W (C1-08105), 2B-81-532Z-W (C1-08107), 2B-81-532Z-W (C1-08108), 2B-81-532Z-W (C1-08109), 2B-81-532Z-W (C1-08110), 2B-81-532Z-W (C1-08119), 2B-81-532Z-W (C1-08129), 2B-81-532Z-W (C1-08130), 2B-81-532Z-W (C1-08131), 2B-81-532Z-W (C1-08132), 2B-81-532Z-W (C1-08093), 2B-81-532Z-W (C1-08094), 2B-81-532Z-W (C1-08102), 2B-81-532Z-W (C1-08111), 2B-81-532Z-W (C1-08122), 2B-81-532Z-W (DP-01372), 2B-81-532Z-W (DP-01434), 2B-81-532Z-W (DP-01596), 2B-81-532Z-W (OR-04381), 2B-81-532Z-W (OR-04382), 2B-81-532Z-W (OR-04405), 2B-81-532Z-W (DP-02792), 2B-81-532Z-W (OR-04579), 2B-81-532Z-W (DP-02933), 2B-81-532Z-W (DP-02940), 2B-81-532Z-W (DP-02949), 2B-81-532Z-W (OR-04600), 2B-81-532Z-W (OR-04625), 2B-81-532Z-W (OR-04609), 2B-81-532Z-W (OR-04709), 2B-81-532Z-W (C1-11368), 2B-81-532Z-W (C1-11369), 2B-81-532Z-W (C1-11370), 2B-81-532Z-W (C1-11371), 2B-81-532Z-W (C1-11374), 2B-81-532Z-W (C1-11375), 2B-81-532Z-W (C1-11376), 2B-81-532Z-W (C1-11377), 2B-81-532Z-W (C1-12452), 2B-81-532Z-W (C1 12452), 2B-81-532Z-W (DP-04710), 2B-81-532Z-W (DP-4763), 2B-81-532Z-W (DP-04778), 2B-81-532Z-W (OR-04842), 2B-81-532Z-W (DP-05418), 2B-81-532Z-W (DP 05418), 2B-81-532Z-W (DP-05899), 2B-81-532Z-W (DP-05967), 2B-81-532Z-W (DP-06761), 2B-81-532Z-W (DP-06985), 2B-81-532Z-W (DP-07012), 2B-81-532Z-W (DP-07791), 2B-81-532Z-W (DP-07792)
TB LEAD, 6F UN-SHROUDED RIGHT HEART CURVE, 020024 - 2B-81-532Z-X (D2-01776), 2B-81-532Z-X (D2-01793), 2B-81-532Z-X (D1-02749), 2B-81-532Z-X (OR-04124), 2B-81-532Z-X (D1-02818), 2B-81-532Z-X (DP-01042), 2B-81-532Z-X (DP-01044), 2B-81-532Z-X (DP-01155), 2B-81-532Z-X (OR-04226), 2B-81-532Z-X (DP 01042), 2B-81-532Z-X (C1-07890), 2B-81-532Z-X (C1-07891), 2B-81-532Z-X (C1-07892), 2B-81-532Z-X (C1-07893), 2B-81-532Z-X (C1-07894), 2B-81-532Z-X (C1-07896), 2B-81-532Z-X (C1-07897), 2B-81-532Z-X (C1-07918), 2B-81-532Z-X (C1-07919), 2B-81-532Z-X (C1-07920), 2B-81-532Z-X (C1-07921), 2B-81-532Z-X (C1-07922), 2B-81-532Z-X (C1-07923), 2B-81-532Z-X (C1-07924), 2B-81-532Z-X (C1-07925), 2B-81-532Z-X (C1-07926), 2B-81-532Z-X (C1-07927), 2B-81-532Z-X (C1-07928), 2B-81-532Z-X (C1-07929), 2B-81-532Z-X (C1-07930), 2B-81-532Z-X (C1-07931), 2B-81-532Z-X (C1-07932), 2B-81-532Z-X (C1-07933), 2B-81-532Z-X (C1-07934), 2B-81-532Z-X (C1-07935), 2B-81-532Z-X (C1-07936), 2B-81-532Z-X (C1-07937), 2B-81-532Z-X (C1-07938), 2B-81-532Z-X (C1-07898), 2B-81-532Z-X (C2-07889), 2B-81-532Z-X (C1-07895), 2B-81-532Z-X (OR-04274), 2B-81-532Z-X (C1-07943), 2B-81-532Z-X (C1-07944), 2B-81-532Z-X (DP-01438), 2B-81-532Z-X (DP-01375), 2B-81-532Z-X (OR-04325), 2B-81-532Z-X (DP-01573), 2B-81-532Z-X (DP-01597), 2B-81-532Z-X (DP-01584), 2B-81-532Z-X (DP-01677), 2B-81-532Z-X (DP-01683), 2B-81-532Z-X (OR-04376), 2B-81-532Z-X (OR-04378), 2B-81-532Z-X (OR-04375), 2B-81-532Z-X (OR-04379), 2B-81-532Z-X (OR-04374), 2B-81-532Z-X (C1-08600), 2B-81-532Z-X (C1-08654), 2B-81-532Z-X (C1-08599), 2B-81-532Z-X (C1-08653), 2B-81-532Z-X (OR-04377), 2B-81-532Z-X (C1-08630), 2B-81-532Z-X (C1-08626), 2B-81-532Z-X (C1-08617), 2B-81-532Z-X (C1-08618), 2B-81-532Z-X (C1-08629), 2B-81-532Z-X (C1-08622), 2B-81-532Z-X (C1-08631), 2B-81-532Z-X (OR-04401), 2B-81-532Z-X (C1-08624), 2B-81-532Z-X (C1-08623), 2B-81-532Z-X (C1-08628), 2B-81-532Z-X (C1-08625), 2B-81-532Z-X (C1-08602), 2B-81-532Z-X (C1-08615), 2B-81-532Z-X (C1-08614), 2B-81-532Z-X (C1-08613), 2B-81-532Z-X (C1-08616), 2B-81-532Z-X (C1-08627), 2B-81-532Z-X (DP-03426), 2B-81-532Z-X (DP-03441), 2B-81-532Z-X (DP-03541), 2B-81-532Z-X (DP-03720), 2B-81-532Z-X (DP-04691), 2B-81-532Z-X (DP 04691), 2B-81-532Z-X (DP-04711), 2B-81-532Z-X (DP-04712), 2B-81-532Z-X (DP-04814), 2B-81-532Z-X (DP-04820), 2B-81-532Z-X (DP-05266), 2B-81-532Z-X (DP-05452), 2B-81-532Z-X (DP-05509), 2B-81-532Z-X (DP-05790), 2B-81-532Z-X (DP-05792), 2B-81-532Z-X (DP-05878), 2B-81-532Z-X (DP-05837), 2B-81-532Z-X (DP-05890), 2B-81-532Z-X (DP-05915), 2B-81-532Z-X (DP-05517), 2B-81-532Z-X (DP-05891), 2B-81-532Z-X (DP-07985)
TB LEAD 4F UNSHROUDED CONVENIENCE KIT, TBK04110USG - 3H-81-003Z-A (DP-03329)
TB LEAD 5F UNSHROUDED CONVENIENCE KIT, TBK05110USG - 3H-81-005Z-A (DP-01161), 3H-81-005Z-A (DP-02004), 3H-81-005Z-A (DP-02009), 3H-81-005Z-A (DP-02435), 3H-81-005Z-A (DP-02884), 3H-81-005Z-A (DP02884), 3H-81-005Z-A (DP 02884), 3H-81-005Z-A (DP-07032)
TB 6F U CONVENIENCE KIT ADELANTE SIGMA PLUS 6F S10CM D15CM, TBK06110USG - 3H-81-007Z-A (D1-02232), 3H-81-007Z-A (D1-02520), 3H-81-007Z-A (D1-02833), 3H-81-007Z-A (DP-01376), 3H-81-007Z-A (DP-02083), 3H-81-007Z-A (DP-02408), 3H-81-007Z-A (DP-02885), 3H-81-007Z-A (DP-03724), 3H-81-007Z-A (DP-04453), 3H-81-007Z-A (DP-05047), 3H-81-007Z-A (DP-06146), 3H-81-007Z-A (DP-06147), 3H-81-007Z-A (OR-04888), 3H-81-007Z-A (DP-07005), 3H-81-007Z-A (DOR05934)
TBRH 4F RT HEART CURVE U CONV. KIT ADELANTE SIGMA PLUS 4F S10CM D15CM, TBRHK04110USG - 3H-81-043Z-A (DP-03330), 3H-81-043Z-A (DP-04258)
TBRH 5F RT HEART CURVE U CONV. KIT ADELANTE SIGMA PLUS 5F S10CM D15CM, TBRHK05110USG - 3H-81-041Z-A (DP-03051)
TBRH 6F RT HEART CURVE U CONV. KIT ADELANTE SIGMA PLUS 6F S10CM D15CM, TBRHK06110USG - 3H-81-034Z-A (FP-054291), 3H-81-034Z-A (FP-54291), 3H-81-034Z-A (C1-08813), 3H-81-034Z-A (DP-02166)
TBV 4F 60 DEGREE U CONV. KIT ADELANTE SIGMA PLUS 4F S10CM D15CM, TBVK04110USG - 3H-81-045Z-A (DP-03397)
TBJ 4F ATRIAL J CURVE U CONV. KIT ADELANTE SIGMA PLUS 4F S10CM D15CM, TBJK04110USG - 3H-81-047Z-A (DP-03396)
Amount recalled
36,688 units

Where it was sold

Why it was recalled

Stop using this product now

During the use of some TB - Temporary Bipolar Pacing Leads, featuring the 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. In some instances, this may cause the wire to be more susceptible to loss of connectivity or breakage during movement of the cables causing interruption of the pacing system.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Oscor, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 4, 2022 · Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.