Reported Nov 5, 2025 · Ongoing
Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322
Greiner Bio-One North America, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Greiner Bio-One North America, Inc. · 4238 Capital Dr · Monroe, NC · 28110-7681
Is this the product you have?
- Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%
- REF454322
- Lot codes and dates
- UDI-DI
29120017574488(Rack)39120017574485(Case)
- LotB250139Q
- Expiration01/09/2026
- UDI-DI
- Amount recalled
- 504,000
Where it was sold
- Nationwide
Why it was recalled
Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections.
This product could cause temporary or reversible health problems — stop using it.