Reported Dec 23, 2015 · Terminated
VentriClear¿ II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF)…
Cook Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cook Inc. · 750 N Daniels Way · Bloomington, IN · 47404-9120
Is this the product you have?
- VentriClear¿ II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF) from the ventricles of the brain.
- Lot codes and dates
- Catalog Number50318Lot
- Numbers
F4365351XF4692818XNS5274254XNS5667436XNS5727900XNS6079165XNS6079166X
- Amount recalled
- 35
Where it was sold
Why it was recalled
Cook Medical initiated a voluntary recall of specific lot numbers of the VentriClear II Ventricular Drainage Catheter Set due to the device being sterilized twice, which may result in the minocycline and rifampin catheter coating to be less than what is indicated on the labeling.
This product could cause temporary or reversible health problems — stop using it.