Reported Dec 14, 2022 · Terminated

LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw…

Integrity Implants Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integrity Implants Inc. · 354 Hiatt Drive · Palm Beach Gardens, FL · 33418

Is this the product you have?

22 of 22 products

Products covered by this recall
#Product as published
15.5mm x 35mm
25.5mm x 40mm
35.5mm x 45mm
45.5mm x 50mm
55.5mm x 55mm
66.5mm x 35mm
76.5mm x 40mm
86.5mm x 45mm
96.5mm x 50mm
106.5mm x 55mm
116.5mm x 60mm
127.5mm x 35mm
137.5mm x 40mm
147.5mm x 45mm
157.5mm x 50mm
167.5mm x 55mm
177.5mm x 60mm
188.5mm x 35mm
198.5mm x 40mm
208.5mm x 45mm
Lot codes and dates

22 of 22 products

Products covered by this recall
Model / numberBarcodesProduct as published
LS-N6SG5535008186130225685.5mm x 35mm, , UDI: (01)00818613022568(10)02210139, lot Number: 2210139.
LS-N6SG5540008186130225755.5mm x 40mm, , UDI: (01)00818613022575(10)02210140, Lot Number: 2210140.
LS-N6SG5545008186130225825.5mm x 45mm, , UDI: ((01)00818613022582(10)02210141, Lot Number: 2210141.
LS-N6SG5550008186130225995.5mm x 50mm, , UDI: (01)00818613022599(10)02210142, Lot Number: 2210142.
LS-N6SG5555008186130226055.5mm x 55mm, , UDI: (01)00818613022605(10)02210143, Lot Number: 2210143.
LS-N6SG6535008186130226436.5mm x 35mm, , UDI: (01)00818613022643(10)02210144, Lot Number: 2210144.
LS-N6SG6540008186130226506.5mm x 40mm, , UDI: (01)00818613022650(10)02210145, Lot Number: 2210145.
LS-N6SG654500818613022667008186130226676.5mm x 45mm, , UDI: (01)00818613022667(10)02210146, Lot Number: 2210146, UDI: (01)00818613022667(10)02210146-1, Lot Number: 02210146-1.
LS-N6SG6550008186130226746.5mm x 50mm, , UDI: (01)00818613022674(10)02210147, Lot Number: 2210147.
LS-N6SG6555008186130226816.5mm x 55mm, , UDI: (01)00818613022681(10)02210148, Lot Number: 2210148.
LS-N6SG6560008186130226986.5mm x 60mm, , UDI: (01)00818613022698(10)02210149, Lot Number: 2210149.
LS-N6SG7535008186130227117.5mm x 35mm, , UDI: (01)00818613022711(10)02210150, Lot Number: 2210150.
LS-N6SG7540008186130227357.5mm x 40mm, , UDI: (01)00818613022735(10)02210152, Lot Number: 2210151.
LS-N6SG7545008186130278917.5mm x 45mm, , UDI: (01)00818613027891(10)02210131, Lot Number: 2210152.
LS-N6SG7550008186130227427.5mm x 50mm, , UDI: (01)00818613022742(10)02210153, Lot Number: 2210153.
LS-N6SG7555008186130227597.5mm x 55mm, , UDI: (01)00818613022759(10)02210154, Lot Number: 2210154.
LS-N6SG7560008186130227667.5mm x 60mm, , UDI: (01)00818613022766(10)02210155, Lot Number: 2210155.
LS-N6SG8535008186130227808.5mm x 35mm, , UDI: (01)00818613022780(10)02210156, Lot Number: 2210156.
LS-N6SG8540008186130227978.5mm x 40mm, , UDI: (01)00818613022797(10)02210157, Lot Number: 2210157.
LS-N6SG8545008186130228038.5mm x 45mm, , UDI: (01)00818613022803(10)02210158, Lot Number: 2210158.
Amount recalled
1,702 screws

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

Why it was recalled

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integrity Implants Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 23, 2023 · Data as of Sep 19, 6:56 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.