Reported Dec 13, 2023 · Ongoing

CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and…

Maquet Medical Systems USA

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Maquet Medical Systems USA · 45 Barbour Pond Dr · Wayne, NJ · 07470-2094

Is this the product you have?

  • CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system.
Lot codes and dates

3 of 3 products

Products covered by this recall
Model / numberBarcodesProduct as published
70104800204037691643526UDI - 04037691643526
701076205Serial Numbers 90425259 to 90425748 (Excluding 90425438, 90425443, 90425568, 90425570, 90425588, 90425716) [***Updated 1/8/24 to include Serial numbers 90421507, 90421980, 90424277. Serial numbers 90425366 and 90423121 are additionally included in the scope due to internal complaints, but these units were corrected prior to the field action initiation and therefore are not included in customer notification***] : No UDI
Serial Numbers 90425438, 90425443, 90425568, 90425570, 90425588, 90425716, 90425747, 90425748
Amount recalled
Updated 1/8/24 - 1858 Total distributed in U.S.; 69 Total affected units in U.S.

Where it was sold

Why it was recalled

Stop using this product now

Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and hypoxia.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Maquet Medical Systems USA recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:40 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.