Reported Dec 23, 2015 · Terminated

MYLA MASTER DVD V4.1 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology…

bioMerieux, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: bioMerieux, Inc. · 100 Rodolphe St · Durham, NC · 27712-9402

Is this the product you have?

  • MYLA MASTER DVD V4.1
  • CLI Product UsageMYLA is a computer application (Middleware) based on Web 2.0
  • technology whichInterfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s))
Lot codes and dates
    • REF # 417191
    • Serial number/Software version V4.1 CLI
Amount recalled
626 systems in total

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

Why it was recalled

MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new bioMerieux, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 4, 2017 · Data as of Sep 19, 5:19 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.