Reported Nov 12, 2025 · Ongoing
Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Beckman Coulter, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter, Inc. · 1000 Lake Hazeltine Dr · Chaska, MN · 55318-1037
Is this the product you have?
- Beckman Coulter UniCel DxI 800
- Part Numbers
973100A71456Photometric Chemistry Analyzer
- Lot codes and dates
973100UDI/DI 15099590265373 (Chaska)
- Serial numbers
302 of 302 codes
610458610455610679610579610432610430610434610664610682610765610500610541610544610545610562610547610585610595610475610453610781610580610444610704610586610684610589610588610493610659610570610498610748610496610560610573610574610577610601610651610680610642610641610683610747610468610772610441610476610489610488610492610532610534610533610535610640610457610464610790
- Serial Numbers
251 of 251 codes
630104630105630160630159630162630110630140630175630143630103630108630174630099630112630113630136630107630102630100630106630115630114630132630111630101630109630146630139630116630126630117630118630120630119630121630122630123630124630125630129630128630130630131630134630135630137630141630142630133630127630164630172630173630144630156630145630157630158630155630154
- Amount recalled
- 657 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.