Reported Nov 18, 2020 · Completed
Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343
Fresenius Medical Care Holdings, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Fresenius Medical Care Holdings, Inc. · 920 Winter St Bld 950 · Waltham, MA · 02451-1521
Is this the product you have?
- Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343
- Lot codes and dates
- All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
- Amount recalled
- 9,907 devices
Where it was sold
- Nationwide
- PR
Why it was recalled
The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Fresenius Medical Care Holdings, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.