Reported Nov 25, 2020 · Terminated
v|tome|x L Series non-destructive testing cabinet x-ray systems
Ge Inspection Technologies LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ge Inspection Technologies Llc · 721 Visions Dr · Skaneateles Falls, NY · 13153-5314
Is this the product you have?
- v|tome|x L Series non-destructive testing cabinet x-ray systems
- Lot codes and dates
v/tome/x L Series (L240L300 or L450)
- Amount recalled
- N/A
Why it was recalled
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.
This product could cause temporary or reversible health problems — stop using it.