Reported Dec 5, 2012 · Terminated
Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow…
Parks Medical Electronics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Parks Medical Electronics, Inc · 19460 SW Shaw St · Aloha, OR · 97007-1242
Is this the product you have?
- Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.
- Lot codes and dates
- Domestic Distribution013053, 013104, 013107, 013124, 013154, 013182, 013193, 014203, 014237, 014248, 014249, 014252, 014253, 014254, 014269, 014276, 014310, 014335, 014339, 014361, 014366, 014376, 014380, 014381, 014382, 014384, 014397, 014401, 014406, 014416, 014424, 014429, 014431, 014443, 014444, 014448, 014454, 014458, 014530, 014532, 014533, 014540, 014542, 014545, 014547, 014551, 014552, 014552, 014553, 014598, 014601, 014604, 014608, 014616, 014654, 014657, 014659, 014660, 014661, 014663, 103077, 105X-600001024, 105X-600001025, 105X-600001026, 105X-600001027, 105X-800001019, 105X-800001022, 105X-800001023, 105X-800001024, 105X-800001026, 105X-800001027, 105X-800001027, 105X-800001028, 105X-800001029, 105X-800001030, 105X-800001031, 105X-E00001001, 105X-E00001002, 105X-E00001003, 105X-E00001004, 105X-E00001005, 105X-E00001007, 105X-E00001008, 105X-E00001009, 105X-E00001010, 105X-E00001011, 105X-E00001012, 105X-E00001013, 105X-E00001014, 105X-E00001016, 105X-E00001017, 105X-E00001018, 105X-E00001019, 105X-E00001020, 105X-F00001000, 105X-F00001001, 105X-F00001002, 105X-F00001003, 105X-F00001004, 105X-F00001005, 105X-F00001005, 105X-F00001006, 105X-F00001007, 105X-F00001008, 105X-F00001009, 105X-F00001010, 105X-F00001011, 105X-F00001012, 105X-F00001013, 105X-F00001014, 105X-F00001015, 105X-F00001016, 105X-F00001017, 105X-F00001018, 105X-F00001020, 105X-F00001021, 105X-F00001022, 105X-F00001023. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify
- Foreign distribution01301, 013043, 013044, 013045, 013046, 013072, 013076, 013175, 013181, 013183, 013188, 014215, 014386, 014392, 014393, 014456, 014535, 014550, 014551, 014561, 014562, 014563, 105X-800001025, 105X-922001017, 105X-922001018, 105X-922001019, 105X-922001021, 105X-9221020, 105X-923001001, 105X-923001002, 105X-923001003, 105X-923001004, 105X-923001005, 105X-923001006, 105X-923001007, 105X-923001007, 105X-923001009, 105X-923001010, 105X-923001011, 105X-E00001000, 105X-E00001003, 105X-E00001006, 105X-E00001015, 105X-F00001019. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify
- Amount recalled
- 349 units (Domestic: 270 units - Foreign: 79 units)
Where it was sold
- Nationwide
Why it was recalled
Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Parks Medical Electronics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.