Reported Jan 12, 2022 · Ongoing

Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A)…

Invacare Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Invacare Corporation · 1200 Taylor St · Elyria, OH · 44035-6248

Is this the product you have?

  • Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers
  • DLX-GRY100-A and DLX-GRY101-A) ModelsTDX
  • SP2
    • TDXSP2
    • TDXSP2-CG
    • TDXSP2X-CG
    • TDXSP2-MCG
    • TDXSP2X-MCG
    • TDXSP2V
    • TDXSP2V-HD
    • TDXSP2HD
    • TDXSP2HD-CG
    • TDXSP2HD-MCG
Lot codes and dates
  • LiNX Gyro firmware version 6.1.2
  • UDI-DI Numbers
    • TDXSP2
    • 00841447100911
    • TDXSP2-CG
    • 00841447100942
    • TDXSP2-MCG
    • 00841447100966
    • TDXSP2HD
    • 00841447108641
    • TDXSP2HD-CG
    • 00841447111207
    • TDXSP2HD-MCG
    • 00841447111214
    • TDXSP2V
    • 00841447101512
    • TDXSP2V-HD
    • 00841447102052
    • TDXSP2X-CG
    • 00841447101000
    • TDXSP2X-MCG
    • 00841447101017
Amount recalled
1477 units US; 1226 units OUS

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Invacare Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.