Reported Dec 12, 2012 · Terminated

The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for…

Natus Medical Incorporated

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Natus Medical Incorporated · 5900 1st Ave S · Seattle, WA · 98108-3248

Is this the product you have?

  • The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling
  • The device is labeled in parts"***OLYMPIC COOL-CAP Cooling Unit***Natus Medical Incorporated 5900 First Avenue South ***Made in USA***Control Unit Model No. 60010***Software in this product is***Technical Service***"
Lot codes and dates
  • 324 of 324 codes

    • 1008
    • 1009
    • 1011
    • 1012
    • 1014
    • 1016
    • 1017
    • 1018
    • 1020
    • 1021
    • 1022
    • 1023
    • 1024
    • 1025
    • 1027
    • 1028
    • 1029
    • 1030
    • 1031
    • 1032
    • 1034
    • 1035
    • 1036
    • 1038
    • 1044
    • 1045
    • 1048
    • 1050
    • 1051
    • 1052
    • 1053
    • 1054
    • 1055
    • 1056
    • 1059
    • 1060
    • 1062
    • 1063
    • 1064
    • 1065
    • 1066
    • 1070
    • 1071
    • 1072
    • 1073
    • 1074
    • 1075
    • 1076
    • 1077
    • 1079
    • 1080
    • 1081
    • 1082
    • 1083
    • 1084
    • 1085
    • 1086
    • 1087
    • 1088
    • 1089
Amount recalled
324 units

Where it was sold

Why it was recalled

Stop using this product now

The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock displayed at the upper right hand corner of the screen stops advancing.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 27, 2015 · Data as of Sep 19, 5:28 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.