Reported Dec 13, 2023 · Ongoing

QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3, Order Number G22655; used to dilate strictures of the biliary tree

Wilson-Cook Medical Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Wilson-Cook Medical Inc. · 4900 Bethania Station Rd · Winston Salem, NC · 27105-1203

Is this the product you have?

  • QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3, Order Number G22655; used to dilate strictures of the biliary tree
Lot codes and dates
    • UDI/DI 00827002226555 (each)
    • 10827002226552 (case)
  • Lot Numbers
    • W4721265
    • W4737428
    • W4740108
    • W4741851
    • W4742110
    • W4742520
    • W4742521
    • W4743198
    • W4744020
    • W4744440
    • W4744875
    • W4749508
    • W4752283
    • W4756468
    • W4720846
    • W4721267
    • W4721796
    • W4722643
    • W4722984
    • W4723305
    • W4723845
    • W4737430
    • W4737767
    • W4738715
    • W4739351
    • W4742522
    • W4742526
    • W4743195
    • W4743196
    • W4744021
    • W4744023
    • W4744443
    • W4744874
    • W4744876
    • W4748041
    • W4748965
    • W4749745
    • W4750014
    • W4750263
    • W4751108
    • W4751109
    • W4751510
    • W4751905
    • W4751907
    • W4753815
    • W4753817
    • W4754120
    • W4754121
    • W4754253
    • W4756473
    • W4758166
    • W4758167
    • W4758929
    • W4758930
    • W4760296
    • W4760714
    • W4761223
    • W4762154
Amount recalled
941 devices

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Why it was recalled

Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Wilson-Cook Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:56 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.