Reported Nov 27, 2024 · Ongoing
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4720
Hobbs Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hobbs Medical, Inc. · 8 Spring St · Stafford Springs, CT · 06076-1505
Is this the product you have?
- Posi-Stop Injection Needle, 23 Gauge
- Catalog Number4720
- Lot codes and dates
- Catalog Number4720
- UDI-DIM84947200
- Lot Numbers
H07-23-103H07-23-104H08-23-020H09-23-111
- Amount recalled
- 70 units
This recall covers 9 products
The agency filed one notice per product. You are reading one of them; the other 8 are below.
Where it was sold
Why it was recalled
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hobbs Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.