Reported Dec 18, 2013 · Terminated
VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is…
Accumetrics Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Accumetrics Inc · 3985 Sorrento Valley Blvd Ste B · San Diego, CA · 92121-1497
Is this the product you have?
- VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide. VerifyNow lIb/IlIa Test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician. The VerifyNow System is a turbidimetric based optical detection system which measures platelet-induced aggregation. The system consists of an instrument, a disposable test device and quality control materials. The VerifyNow lIb/IlIa Test is a semi-quantitative, whole blood platelet function test used to measure glycoprotein (GP) lIb/IlIa receptor blockade in patients treated with abciximab or eptifibatide.
- Lot codes and dates
Lot No. WC0181HWC0181JWC0181K
- Amount recalled
- 69 units
Where it was sold
Why it was recalled
Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the test within 15 minutes of collecting the patient sample.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Accumetrics Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.