Reported Dec 3, 2014 · Terminated
Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators…
Respironics California Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Respironics California Inc · 2271 Cosmos Ct · Carlsbad, CA · 92011-1517
Is this the product you have?
- Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators are microprocessor-controlled, electrically powered mechanical ventilators. Intended for use by medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients.
- Lot codes and dates
- All Esprit V1000 ventilators manufactured prior to December 31, 2004 and those units manufactured prior to 12/2004 that have been converted to V200. Serial NumbersVSTEST
641 of 641 codes
VS3000584VS3000614VS3001023VS3001026VS3001046VS3001122VS3001123VS3001643VS3001814VS3001815VS3002027VS3002079VS3004925VS3001953VS3001979VS3002122VS3004232VS3004312VS3004336VS3004342VS3004348VS3004946VS3005661VS3005779VS3000536VS3000538VS3000542VS3000927VS3000935VS3001253VS3001255VS3001314VS3001355VS3001388VS3001986VS3002078VS3002625VS3002737VS3002768VS3002804VS3002832VS3002835VS3003127VS3003194VS3003199VS3003218VS3004609VS3004841VS3000570VS3000638VS3001007VS3001017VS3001027VS3001110VS3001222VS3001256VS3001266VS3001313VS3001501VS3001502
2,307 of 2,307 codes
VS3004712VS3001895VS3001930VS3002806VS3003633VS3005851VS3005915VS3005982VS3005985VS3006071VS3006185VS3000560VS3000567VS3000575VS3000602VS3000607VS3000617VS3000635VS3000906VS3000907VS3000909VS3000911VS3000912VS3000939VS3000943VS3000946VS3000952VS3000957VS3000960VS3000964VS3000969VS3000970VS3000973VS3001019VS3001032VS3001035VS3001036VS3001048VS3001059VS3001060VS3001061VS3001065VS3001066VS3001067VS3001068VS3001070VS3001074VS3001075VS3001076VS3001078VS3001084VS3001087VS3001094VS3001103VS3001109VS3001111VS3001113VS3001117VS3001630VS3001631
- All Esprit V1000 ventilators manufactured prior to December 31, 2004 and those units manufactured prior to 12/2004 that have been converted to V200. Serial Numbers
- Amount recalled
- 3,943 units total (3,071 units in US)
Where it was sold
- Nationwide
Why it was recalled
If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Respironics California Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.