Reported Jan 19, 2022 · Terminated
LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
Covidien LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Covidien Llc · 15 Hampshire St · Mansfield, MA · 02048-1113
Is this the product you have?
- LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
- Lot codes and dates
- UDI20884521705880 10884521705883
- Lot Number11320187X
- Amount recalled
- 504 devices
Why it was recalled
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.