Reported Dec 19, 2012 · Terminated
Medtronic SynchroMed II, Model 8637, (The SynchroMed¿ II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized…
Medtronic Neuromodulation
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Neuromodulation · 7000 Central Ave NE · Minneapolis, MN · 55432-3568
Is this the product you have?
- Medtronic SynchroMed II, Model 8637, (The SynchroMed¿ II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II Infusion System include the pump, catheter, and catheter accessories.
- Lot codes and dates
- All SynchroMed EL and SynchroMed II pumps
- Amount recalled
- 174,700 devices total
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Stop using this product now
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.