Reported Dec 19, 2012 · Terminated
Reusable, Self-Adhering TENS/NMES/FES electrodes Single Patient Use Only. An electrocardiograph electrode is the…
Covidien LLC dba Uni-Patch
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Covidien LLC dba Uni-Patch · 1313 Grant Blvd W · Wabasha, MN · 55981-1058
Is this the product you have?
- Reusable, Self-Adhering TENS/NMES/FES electrodes Single Patient Use Only. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.
- Lot codes and dates
Lots 202404202506203431
- Amount recalled
- 3594 pouches
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Covidien is conducting a voluntary recall regarding specific lots of TENS electrodes because it was determined that during the manufacturing process the surface of the wires in these lots may have been exposed to an employee who had a minor cut.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Covidien LLC dba Uni-Patch recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.