Reported Jan 26, 2022 · Terminated
Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Draeger Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042
Is this the product you have?
- Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
- Lot codes and dates
- UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002
- Amount recalled
- 5
Where it was sold
Why it was recalled
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
This product could cause temporary or reversible health problems — stop using it.