Reported Feb 14, 2018 · Terminated

The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are…

Microline Surgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Microline Surgical, Inc. · 800 Cummings Ctr Ste 166T · Beverly, MA · 01915-6171

Is this the product you have?

  • The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are indicated for cutting, grasping, dissecting, and coagulation of tissue in endoscopic and laparoscopic surgical procedures.
Lot codes and dates
  • All lots of product codes
    • 3201
    • 3221
    • 3231
    • 3241
    • 3251
    • 3261
    • 3271
    • 3281
    • 3291
    • 3301
    • 3311
    • 3321
    • 3331
    • 3341
    • 3351
    • 3361
    • 3371
    • 3381
    • 3401
    • 3411
    • 3421
    • 3431
  • UDI's for listed lots of product codes
    • 3201-UDI 00811099011644
    • 3221-UDI 00811099011651
    • 3231-UDI 00811099011668
    • 3241-UDI 00811099011675
    • 3251-UDI 00811099011682
    • 3261-UDI 00811099011699
    • 3271-UDI 00811099011705
    • 3281-UDI 00811099011712
    • 3291-UDI 00811099011729
    • 3301-UDI 00811099011859
    • 3311-UDI 00811099011736
    • 3321-UDI 00811099011743
    • 3331-UDI 00811099011767
    • 3341-UDI 00811099011774
    • 3351-UDI 00811099011781
    • 3371-UDI 00811099011804
    • 3381-UDI 00811099011811
    • 3401-UDI 00811099011828
    • 3411-UDI 00811099011835
    • 3421-UDI 00811099011842
    • 3431-UDI 00811099011644
Amount recalled
61,769 units

Where it was sold

  • Nationwide

Why it was recalled

The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off during a surgical procedure and if cautery is needed, there is a potential for burn due to insulation failure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Microline Surgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 6, 2020 · Data as of Sep 19, 4:45 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.