Reported Dec 26, 2012 · Terminated

Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable…

Bunnell, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Bunnell, Inc. · 436 Lawndale Dr · Salt Lake City, UT · 84115-2917

Is this the product you have?

  • Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters
  • Labeling states manufactured byBunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use
Lot codes and dates
  • Patient Circuit (catalog # 902) lot # on
  • each individual patient circuit label12C092, 12C102, 12C115, 12C125, 12C136, 12D159, 12D172, 12D189, 12E204, 12E211, 12E224, 12F241, 12F254, 12F271, 12G279, 12G290, 12G307, 12G321, 12H330, 12H349, 12I362, 12I371, 12I397, 12J413, 12J430, 12J448, 12K457, 12K471 Patient Circuit Kit (catalog # 397) lot #
  • on the label on outside of box
    • 12C099
    • 12C114
    • 12C124
    • 12C135
    • 12D152
    • 12D170
    • 12D188
    • 12E203
    • 12E210
    • 12E223
    • 12F238
    • 12F255
    • 12F274
    • 12G209
    • 12G280
    • 12H334
    • 12H341
    • 12I356
    • 12I369
    • 12I392
    • 12J411
    • 12J425
    • 12J447
    • 12K467
    • 12K479
Amount recalled
5743 patient circuits distributed in commerce

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Bunnell, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 21, 2014 · Data as of Sep 19, 6:17 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.