Reported Dec 19, 2012 · Terminated

Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitro measurement of pH,…

Roche Diagnostics Operations, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Roche Diagnostics Operations, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025

Is this the product you have?

  • Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Na+ K+, iCa2+ (ISE), hematocrit (Hct), metabolites (Gluc, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHB, COHB, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory.
Lot codes and dates
  • Part number
    • 05169992001 cobas b 123 Fluid Pack COOX
    • 200 with Lot# 21426121 and Lot# 21426171
Amount recalled
23 units of Fluid Pack COOX, 200 were distributed

Where it was sold

Why it was recalled

It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a change in the thickness in the cuvette layer. Internal investigations found the altered thickness of the cuvette was due to production equipment settings. Cuvettes were produced within specifications, but within the upper

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new produce recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 7, 2013 · Data as of Sep 19, 5:30 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.