Reported Dec 25, 2013 · Terminated
Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1,…
Karl Storz Endoscopy America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Karl Storz Endoscopy America Inc · 15 Wells St · Southbridge, MA · 01550-4503
Is this the product you have?
- Intubation Flexible Fiberscope
- Models11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. " Transoral intubation " Transnasal intubation " Evaluation of placement of endotracheal and double lumen endotracheal tubes " Provide visual access to the larynx and tracheobronchial tree
- Lot codes and dates
C1140C1140EC1160C1160EQKC1750QKC1750EQKC1694QKC1694E
- Amount recalled
- 1516 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Karl Storz Endoscopy America Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.