Reported Dec 4, 2024 · Ongoing
smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
Smiths Medical ASD, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Smiths Medical ASD, Inc. · 6000 Nathan Ln N · Minneapolis, MN · 55442-1690
Is this the product you have?
- smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
- Lot codes and dates
REF 14-504-17UDI/DI 35021239002579
- Lot Numbers
421 of 421 codes
000334000036200003616000358000035420003509000357900035680003599000361500035890003562000360100036180003583000361300032700003629000329300035840003487000361700032800003348000347000035230003607000357600034220003614000363300035780003630000320700034380003559000357000034780003533000358800034330003556000356100034460003284000354500034610003590000339800035300003539000355000032130003598000362300035040003548000355200036310003594
- Lot Numbers
000321500034150003483
- Amount recalled
- 142327 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Why it was recalled
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Smiths Medical ASD, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.