Reported Dec 16, 2020 · Terminated

Philips IntelliVue MP2 Patient Monitors (865040) - Product Usage: intended for monitoring and recording of, and to…

Philips North America LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America Llc · 222 Jacobs St · Cambridge, MA · 02141-2289

Is this the product you have?

  • Philips IntelliVue MP2 Patient
  • Monitors (865040) - Product Usageintended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of the hospital environment
Lot codes and dates
  • Serial NumbersDE73602213
  • DE73602577 UDI(01)00884838000278(21)DE73602213 (01)00884838000278(21)DE73602577
Amount recalled
2 units OUS

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • SD

Why it was recalled

Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 22, 2021 · Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.