Reported Dec 12, 2018 · Terminated

Captus 3000 Thyroid Uptake System, Model # 5430-0076 and 5430-0077

Capintec Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Capintec Inc · 7 Vreeland Road · Florham Park, NJ · 07932

Is this the product you have?

    • Captus 3000 Thyroid Uptake System
    • Model # 5430-0076 and 5430-0077
Lot codes and dates
  • Serial Numbers 901100 through 901309. Serial Numbers 901310 through 901965 as of 1/15/19.
Amount recalled
385

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Capintec Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 30, 2020 · Data as of Sep 19, 6:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.