Reported Feb 14, 2018 · Terminated

Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of…

The Binding Site Group, Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: The Binding Site Group, Ltd. · 8 Calthorpe Road · Birmingham, N/A · N/A

Is this the product you have?

  • Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of rheumatoid factor in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings
Lot codes and dates
    • Lot #416595
    • Exp. 31/03/2018
    • UDI #05051700018319
Amount recalled
49 kits to the U.S. office (32 kits further distributed)

Where it was sold

Why it was recalled

The low and high controls for the product are returning results above the assigned values.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new The Binding Site Group, Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 9, 2020 · Data as of Sep 19, 6:13 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.