Reported Dec 4, 2024 · Ongoing

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Shimadzu Corp. Analytical & Measuring Instruments Div.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Shimadzu Corp. Analytical & Measuring Instruments Div. · 1 Chome 1 · Kyoto · N/A

Is this the product you have?

  • Autosampler
  • Models
    • SIL-40C XR CL
    • SIL-40C X3 CL
Lot codes and dates
  • Model/UDI-DI/Serial NumbersSIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16
Amount recalled
48

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Shimadzu Corp. Analytical & Measuring… recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:41 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.