Reported Jan 13, 2016 · Terminated

CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic…

Partners In Medicine LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Partners In Medicine Llc · 11469 Olive Blvd Ste 127 · Saint Louis, MO · 63141-7108

Is this the product you have?

  • CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic parameters time synchronized using Impedance Cardiography (ICG), Electrocardiography (ECG), oscillometric Blood Pressure (oscBP) and continuous Blood Pressure (contBP). Furthermore Sympathic Tone and Parasympathic Tone are calculated by power spectral analysis of ECG (heart rate variability) and continuous blood pressure (blood pressure variability) as well as the Baroreceptor Reflex Sensitivity (BRRS). For use by medically trained personnel. It has been designed for diagnoses aiding and must not be used for vital sign monitoring of critically ill patients under absence of additional, suitable self-monitoring devices or medical trained personnel. The device measures continuously the subject's hemodynamic parameters without reporting any diagnosis. It is the physician's responsibility to make proper judgment based on these parameters.
Lot codes and dates
  • Serial numbers
    • 003040i-2014-161-014-GG-0000
    • 003040i-2015-101-003-GG-0000
    • 003040i-2014-161-001-GG-0000
    • 003040i-2014-161-002-GG-0000
    • 003040i-2014-161-003-GG-0000
    • 003040i-2014-161-004-GG-0000
    • 003040i-2014-161-005-GG-0000
    • 003040i-2014-161-006-GG-0000
    • 003040i-2014-161-007-GG-0000
    • 003040i-2014-161-008-GG-0000
    • 003040i-2014-161-009-GG-0000
    • 003040i-2014-161-010-GG-0000
    • 003040i-2014-161-011-GG-0000
    • 003040i-2014-161-012-GG-0000
    • 003040i-2014-161-013-GG-0000
    • 003040i-2014-161-015-GG-0000
    • 003040i-2014-161-016-GG-0000
    • 003040i-2014-161-017-GG-0000
    • 003040i-2014-161-018-GG-0000
    • 003040i-2014-161-019-GG-0000
    • 003040i-2014-161-020-GG-0000
    • 003040i-2014-161-021-GG-0000
    • 003040i-2014-161-022-GG-0000
    • 003040i-2014-161-023-GG-0000
    • 003040i-2014-161-024-GG-0000
    • 003040i-2014-161-025-GG-0000
    • 003040i-2015-101-001-GG-0000
    • and 003040i-2015-101-002-GG-0000
Amount recalled
28 worldwide 2 monitors distributed in the U.S.

Where it was sold

Why it was recalled

Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Partners In Medicine LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 17, 2017 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.