Reported Dec 27, 2023 · Ongoing
Injection Needle, LUER-lock,0.8mm, REF: 26175PD
Karl Storz Endoscopy
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Karl Storz Endoscopy · 2151 E Grand Ave · El Segundo, CA · 90245-5017
Is this the product you have?
- Injection Needle, LUER-lock,0.8mm
- REF26175PD
- Lot codes and dates
- UDI-DI04048551099391. All lot numbers
- Amount recalled
- 1,127
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
Why it was recalled
The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.