Reported Jan 13, 2016 · Terminated

CONTENDER POST-OP-KNEE BRACE FULL FOAM Model: 75-7550-000 Post-Operative Knee Brace is indicated for range of motion…

Corflex

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Corflex · 669 E Ind Park Dr · Manchester, NH · 03109-5625

Is this the product you have?

  • CONTENDER
  • POST-OP-KNEE BRACE FULL FOAM Model75-7550-000 Post-Operative Knee Brace is indicated for range of motion control at the knee joint following ACL, PCL, MCL, and LCL surgeries, meniscal repairs, patella realignment, regenerative chondroplasty, stable femoral fractures, total knee replacements, and high tibial osteotomies
Lot codes and dates
  • (1) all braces that begin with a Revision Code of AN, AP, AQ, AR, AS, AT, AU are to be returned to Corflex immediately for review; (2) all braces with a Revision Code beginning with
  • AM and that comply with the date range
    • Braces manufactured between September 1
    • 2015 and November 6
    • 2015
Amount recalled
80 units

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide

Why it was recalled

Knee brace could fail to maintain the locked position and a loss of balance or undesired flexion/extension may occur at the affected knee joint,

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Corflex recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 29, 2016 · Data as of Sep 19, 6:01 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.